Skip to main content

Table 4 Fraction of MPET and expected Raman signal reached by some types of lasers

From: Optimization of laser wavelength, power and pulse duration for eye-safe Raman spectroscopy

Laser characteristics

Fraction of MPET (%)

Expected Raman signal (a.u.)

Laser type

wl

power

PRR

E0

t

T (s)

T (s)

 

nm

W

Hz

mJ

ns

0.01

0.1

1

10

0.01

0.1

1

10

355

0.2

50

4

11

1

6

19

61

0.28

2.69

15.1

85.0

Lamp pumped, tripled Nd:YAG

355

0.7

100

7

12

3

10

33

100

0.82

4.62

26.0

140

Diode pumped, tripled Nd:YAG

355

4

4 104

0.1

2

23

100

100

100

5.61

44.2

78.5

140

Diode pumped, tripled Nd:YAG

355

8

5 104

0.16

25

45

100

100

100

11.2

44.2

78.5

140

Diode pumped, tripled Nd:YAG

351

15

750

20

18

17

52

100

100

4.26

24.0

81.2

144

Excimer, XeFl

266

0.08

20

4

11

100

100

100

100

1.00

1.00

1.00

1.00

Lamp pumped, quadrupled Nd:YAG

405

0.5

CW

500

109

100

100

100

100

0.01

0.03

0.17

0.96

CW diode laser

325

0.1

CW

100

109

1

3

18

100

0.18

1.83

18.3

182

He-Cd CW laser

1570

1.5

30

50

6

2

15

100

100

0.02

0.24

1.62

1.62

OPO-shifted Er:glass